Using medical pro, Proceed as follows – Barco DICOM THEATER LITE MGP D5 User Manual

Page 77

Advertising
background image

Using MediCal Pro

78

MGP D5 Dicom Theater Lite

Using MediCal Pro

After installation of the hardware and software, you must start
MediCal Pro to calibrate the display(s).

Proceed as follows:

1. Start MediCal Pro. If appropriate, you can connect to MediCal

Administrator.

2. During startup, MediCal Pro will detect the connected review

display(s). They will be added automatically to the system
configuration in MediCal Pro.

Note: In case the review displays were not automatically
added to the configuration, you can add them manually, by
running the Configuration Setup Wizard.

3. For all the devices in the configuration, check if the DPMS

setting is turned on. Therefore, right-click on the display icon
and select Properties... from the drop-down menu. Then click
on Details...

Check if the Powersave option is checked. If not, check it.
This is necessary to use the DPMS possibilities of the imaging
board.

4. For all the devices in the configuration, define (if necessary)

and select a Preset.

5. After selecting the Preset, MediCal Pro starts consistency

calibration automatically.

Please refer to the MediCal Pro User Guide for more information.

Advertising