To calibrate a display using the barco lcd sensor, Click the calibrate button – Barco DICOM THEATER LITE MGP D5 User Manual

Page 90

Advertising
background image

Using TheaterWatch

MGP D5 Dicom Theater Lite

91

3. A wizard starts, guiding you through the calibration

procedure. Follow the guidelines in the wizard.

The projector calibration procedure is a combination of
actions in TheaterWatch and projector OSD (on-screen
display) manipulations.

To calibrate a display using the Barco LCD sensor:

1. If you have multiple displays of the same type in the system,

consider if you wish to equalize the display you calibrate to
a reference display.

Equalization means the luminance of the display you are
calibrating will be matched to the luminance of the reference
display.

If you wish to equalize the display, you must close the
Display Settings dialog and make sure the Equalization
option in the Application Settings dialog is set. See
“Calibration tab” in the description of the Application
Settings dialog.

2. If you have multiple projectors or displays connected (multi-

head system), select the display you wish to calibrate by
clicking on the corresponding numbered icon in the Display
Settings dialog.

3. Click the Calibrate button.

4. A message appears, showing the different calibration

options.

Figure 34: Calibration choices