ISMATEC SA/Ecofluidic/15.03.07/CB/GP

20

Status Indication

LED

Operation

Status

class

Cause

Solution

green

Pump runs normally

normal

No fault

yellow

Pump stand-by

normal

No fault

flashes

green-yellow

Adaptive speed ramp-up

Info

The set speed requires more torque than available. Thus, the actual speed is

adaptively ramped up (duration: max. 5 min).

1 The backpressure is too high.

2 Under the application conditions, peak pressures occur during start-up.

3 A new (hard!) tube is being used.

4 The tubing had been fully squeezed for a long time and suffers deformation.

5 Tubing wall thickness is larger than 1.6 mm (1/16 inch).

6 A high viscosity medium is being pumped.

7 Mains voltage is less than nominal - 10%.

Let the pump continue to run! If it shuts down with a red LED: check operation conditions.

flashes

red-yellow

Temperature warning

Info

The internal temperature of the pump is becoming critical.

1 The fan is broken.

2 The environmental temperature is above 40°C (104°F).

1 Check if the fan expells air through the ventilation slots on the side. If it doesn‘t, contact your

ISMATEC service center.

2 Provide for enough ventilation around the pump.

flashes green

Undervoltage

Info

Mains voltage is less than nominal - 10%.

Check the mains supply.

flashes yellow Pumphead opened up while

running

Fault

The pumphead was opened while the pump was running.

1 For safety reasons, the lid should be opened only when the pump is at rest.

2 The closed lid is not perfectly in position because the tubing is missing.

1 Close the lid and press the control knob once. The pump switches to ‚stand-by‘.

2 Insert a tube and close the lid. Press the control knob once to switch the pump to ‚stand-by‘.

flashes red

Overtemperature shut-down Fault

Due to overheating, the drive shut down automatically.

1 The fan is broken.

2 The ventilation slots are blocked.

3 The environmental temperature is above 40°C.

Press the control knob once. If the LED changes to yellow (‚stand-by‘), the pump has cooled down

already. If not, wait for a couple of minutes.

In any case, check if:

1 the fan expells air through the slots on the side. If it doesn‘t, contact you ISMATEC service

center.

2 the ventilation slots on the side and in the bottom are blocked. Clean them if necessary.

3 the environmental temperature is above 40°C.

red

Overload shut-down

Fault

Maximum torque has been exceeded:

1 The drive is blocked.

2 An unsuitable tubing was inserted.

3 The tubing is not mounted correctly.

4 System pressure has changed and exceeds the specified maximum

differential pressure.

5 Viscosity has gone up.

6 Mains voltage is less than nominal - 10%.

1 Make sure that the pumphead is not blocked.

2 Only suitable peristaltic tubing must be used (diameter, material, wall thickness - see page 22).

3 Ascertain that the tubing has been properly inserted (see page 13).

4 Verify that no excess backpressure or vacuum has build up at the outlet or inlet (e.g. by

congestion, valves, filters, empty reservoirs without venting).

5 Inspect the medium for its temperature or other physical or chemical changes.

6 Check the mains supply.

Press the control knob once to reset the pump. Its LED changes to stand-by

flashes yellow

after power-on

Transmission of diagnostic

data

normal

Right after power-up, the pump sends service information through the LED for

service purposes.

No fault