BNC 845-M User Manual

Page 32

Advertising
background image

 

32 
 

Model 845-M Programmer’s Manual V1.1 2011


[SOURce]:LFOutput Subsystem

Command

Parameters

Unit

(default)

[SOURce]:LFOutput:AMPLitude

<value>

1 V

[SOURce]:LFOutput:FREQuency

<value>

400 Hz

[SOURce]:LFOutput:STATe

ON|OFF|1|0

OFF

[SOURce]:LFOutput:SHAPe

SINE|TRIangle|SQUare

SINE

[SOURce]:LFOutput:SOURce

LFGenerator|PULM|TRIGger

LFG

:LFOutput:AMPLitude
[SOURce]:LFOutput:AMPLitude <value>
[SOURce]:LFOutput:AMPLitude?

This command sets the low frequency generator amplitude. This setting does only take effect if

:LFOutput:SOURce is set to LFGenerator and LFOutput:SHAPe is either set to SINE or TRIangle. Using
any other setting, the output amplitude is fixed 2.5 V.

The variable <value> is expressed in units of Volt.
Range 0-2.5

*RST 1

:LFOutput:FREQuency
[SOURce]:LFOutput:FREQuency <value>
[SOURce]:LFOutput:FREQuency?

This command sets the low frequency generator frequency. This setting does only take effect if

:LFOutput:SOURce is set to LFGenerator.

The variable <value> is expressed in units of Hertz.
Range 10-5000000

*RST 400

:LFOutput:STATe
[SOURce]:LFOutput:STATe ON|OFF|1|0
[SOURce]:LFOutput:STATe?

This command sets the function output / low frequency generator state.

*RST OFF

:LFOutput:SHAPe

[SOURce]:LFOutput:SHAPe SINE|TRIangle|SQUare

[SOURce]:LFOutput:SHAPe?

This command sets the low frequency generator waveform. This setting does only take effect if

:LFOutput:SOURce is set to LFGenerator.

SINE Sine wave output, amplitude is selectable by the :LFOutput:AMPLitude setting.
TRIangle Triangle wave output, amplitude is selectable by the :LFOutput:AMPLitude

setting.

SQUare Square wave output, amplitude is fixed 2.5 V.

*RST SINE

:LFOutput:SOURce

[SOURce]:LFOutput:SOURce LFGenerator|PULM|TRIGger

[SOURce]:LFOutput:SOURce?

Advertising