5 Maintenance

23

Eppendorf Research® plus — Operating manual

The density of distilled water is approx. 0.9982 mg/μL at 20°C and 0.9965 mg/μL at 27°C.
If the set volume corresponds to the actual volume, no correction is necessary.
If there is a difference between the actual volume and the set volume of distilled water, please

check the following:

Is there any liquid dripping from the tip?

Is the pipette tip fitted leak-proof?

Is the tip cone undamaged?

Are the piston and the cylinder leak-proof?
Adequate leak tightness is ensured when no drop is formed at the pipette tip after aspiration

of the nominal volume with distilled water and a waiting time of approx. 15 s. Hold the pipette

vertically, making sure not to touch the pipette tip. Prewet the tip several times in the case of

nominal volumes ≤ 20 μL.

Does the temperature of the pipetted liquid correspond to:

– the temperature of the device?
– the ambient air temperature?

Is the weighing location free from drafts?

Does the work method and pipetting speed permit complete aspiration and dispensing of the

liquid?

Has the correct numerical value for "Density liquids at weighing temperature" been used for

the calculation of the actual volume?

Is the set volume correct?

For very small volumes (<10 μL): is the fine balance sufficiently sensitive (balance resolution:

0.001 mg)?

Were original epT.I.P.S pipette tips used for testing?
Information as to which pipette tip must be used as a test tip for checking the technical data

can be found in the "Technical Data" chapter of the operating manual.

No adjustment changes are allowed unless you can answer all the questions with "yes". In all

other cases, the problems associated with the questions answered with "no" must be eliminated.

If the problem is remedied by exchanging a complete lower part or other parts that have an effect

on the volume, proper assembly must be verified by carrying out a gravimetric test. Information

on the systematic and random errors to be met can be found in the "Technical Data" chapter.