2 Safety

9

Eppendorf Research® plus — Operating manual

NOTICE!

NOTICE! Carry-over, contamination and incorrect dispensing results due to the

incorrect use of pipette tips.
The pipette tips are for single use only. Prolonged use can have a negative impact on

dispensing tasks.



Use the pipette tips only once.



Do not autoclave the epT.I.P.S. Dualfilter.

NOTICE!

NOTICE! Incorrect dispensing volume for special liquids and from temperature

differences.
Solutions which differ greatly from water in terms of their physical data, or temperature

differences between the pipette, pipette tip and liquid can result in incorrect dispensing

volumes.



Avoid temperature differences between pipette, pipette tip and liquid.



Make sure that the temperature is constant, between 20 and 25°C and at ±0.5°C.



Check the dispensing volume and readjust the pipette in case of deviations.