Eppendorf Research® plus — Operating manual

30 - 300 μL
Increment:
0.02 μL

orange
20 - 300 μL
55 mm

30 μL

± 13.0

± 3.9

± 5.0

± 1.5

150 μL

± 3.0

± 3.9

± 1.0

± 1.5

300 μL

± 1.3

± 3.9

± 0.5

± 1.5

100 - 1 000 μL
Increment:
1 μL

blue
50 - 1 000 μL
71 mm

100 μL

± 8.0

± 8.0

± 3.0

± 3.0

500 μL

± 2.0

± 8.0

± 0.6

± 3.0

1 000 μL

± 0.8

± 8.0

± 0.3

± 3.0

0.5 - 5 mL
Increment:
5 μL

purple
0.1 - 5 mL
120 mm

0.5 mL

± 8.0

± 40.0

± 3.0

± 15.0

2.5 mL

± 1.6

± 40.0

± 0.6

± 15.0

5.0 mL

± 0.8

± 40.0

± 0.3

± 15.0

1 - 10 mL
Increment:
10 μL

turquoise
1 – 10 ml
165 mm

1.0 mL

± 6.0

± 60.0

± 3.0

± 30.0

5.0 mL

± 1.2

± 60.0

± 0.6

± 30.0

10.0 mL

± 0.6

± 60.0

± 0.3

± 30.0

Error limits in accordance with ISO 8655-2 - Research plus variable single-channel
Model

Test tip
epT.I.P.S. Color

code
Volume range
Length

Testing

volume

Error limits

Error

Systematic

Random

± %

± μL

± %

± μL