2 research plus multi-channel variable, 6 technical data – Eppendorf Research plus User Manual

Page 37

Advertising
background image

6 Technical data

37

Eppendorf Research® plus — Operating manual

6.2 Research plus multi-channel variable

Research plus variable multi-channel
Model

Test tip
epT.I.P.S. Color

code
Volume range
Length

Testing

volume

Error limits

Error

Systematic

Random

± %

± μL

± %

± μL

0.5 – 10 μL
Increment:
0.01 μL

medium gray
0.1 - 20 μL
40 mm

0.5 μL

± 12

± 0.06

± 8.0

± 0.04

1 μL

± 8.0

± 0.08

± 5.0

± 0.05

5 μL

± 4.0

± 0.2

± 2.0

± 0.1

10 μL

± 2.0

± 0.2

± 1.0

± 0.1

10 – 100 μL
Increment:
0.1 μL

yellow
2 - 200 μL
53 mm

10 μL

± 3.0

± 0.3

± 2.0

± 0.2

50 μL

± 1.0

± 0.5

± 0.8

± 0.4

100 μL

± 0.8

± 0.8

± 0.3

± 0.3

30 – 300 μL
Increment:
0.2 μL

orange
20 - 300 μL
55 mm

30 μL

± 3.0

± 0.9

± 1.0

± 0.3

150 μL

± 1.0

± 1.5

± 0.5

± 0.75

300 μL

± 0.6

± 1.8

± 0.3

± 0.9