5 Maintenance

29

Eppendorf Research® plus — Operating manual

The values mentioned are theoretical values and are for orientation purposes only. The

volume changes mentioned apply to each volume setting. For the volume change, you

should first set the optimal value for 10% of the nominal volume and then carry out a

gravimetric test. Afterwards carry out gravimetric tests to check 50% and 100% of the

nominal volume with this setting. Change the selected setting again if necessary to achieve

the best possible correction for all volumes. Use the error limits in accordance with ISO

8655-2 (see p. 31) and the technical data provided by Eppendorf AG (see p. 35) to decide

whether the data obtained meets your requirements.

7. If the measured gravimetric values meet your requirements: Slide the locking mechanism up.

Multi-channel
Nominal volume
Color code

+½ revolution

+¼ revolution

–¼ revolution

–½ revolution

10 μL
medium gray

0.53 μl

0.27 μl

–0.27 μL

–0.53 μL

100 μL
yellow

5.4 μl

2.7 μl

–2.7 μL

–5.4 μL

300 μL
orange

10.7 μl

5.4 μl

–5.4 μL

–10.7 μL

6. Slide the locking mechanism up and carry

out a gravimetric test of the changes which

have been made.

5

Main

tenan

ce