3.4 Storage

Before storing the Clean Air Flow

TM

system, it should be thoroughly dry.

Do Not Keep used fi lter in the same enclosed storage space as new fi lters.

Store the system in a clean area, away from moisture, heat and direct

sunlight. Storage temperatures should not exceed 120

0

F.

3.3 Inspection

1. After cleaning (section 3.2), inspect the individual parts and ensure that all
components and connection sites are clean and in good condition.
2. Inspect for damages; such as, holes, cracks, and scratches on all parts. If in
doubt of effective respiratory performance replace the parts with only ArcOne ap-
proved parts for this system.
3. Additionally, inspect the following:

a. Blower unit: Button function. Replace if damaged.

b. Breathing Tube:

o O-ring state. Replace if worn/cut, torn, or if in doubt as to

condition.

o Check for holes or splitting. When in doubt, replace the

Breathing Tube. NEVER patch the Breathing Tube.

c. Faceseal: ensure that the face seal has not lost its elasticity and that

there are no tears or holes. Replace if in doubt as to the condition of

the face seal.

d. Waist belt: ensure the clip securely snaps together.

4.1 Battery

Refer to fi g. 15 for the following steps.
Removal:

1. Pull up on the snap tab.

2. While holding the snap tab away
from the blower pull the battery in
the direction indicated.
Installation: See section 2.1

4.0 Component Replacement

Snap pull

Remove

Direction

#1

Battery

Blower

Figure 15