Ramp dwell, Setting up a ramp dwell program, 6 ramp dwell programming – Carbolite 301 User Manual

Page 13

Advertising
background image

Ramp Dwell Programming

13

MC16-GB-C

– 1.04

6 Ramp Dwell Programming


The 301 controller has the capability to follow a Ramp Dwell program.
















6.1

Setting up a Ramp Dwell program


Set the Controller to Hold Mode:



Start at the Home display
Press and hold the Up and Down arrow keys together for 1.5 seconds.
The display will flash

HOLd to show that hold mode has been entered.



Set the Timer Type to

t1





Start at the Home display.
Press and hold the Page key for 1.5 seconds to enter the set up menu.
Press the Page key until

t.tYP shows on the display.

Use the Up and Down arrow keys to set the value to

t1

The value will then be stored without any further input.

See Section 5.6 for the timer temperature band setting.

Time

Setpoint

ramp rate

SPrr

Dwell time

t1

T

em

pe

ra

ture

Natural cool

down

Hold Together

t.tYP

Hold