Advanced Operation

9

MC16-GB-C

– 1.04

3.4

Changing the Maximum Output Power


Note: Output Power is a product specific setting and will not appear on all furnaces and
ovens

See section 7 for more information on Output Power.





Start at the Home display.
Press and hold the Page key for 1.5 seconds to enter the set up menu.
Press the Page key until

OPHi shows on the display.

Use the Up and Down arrow keys to alter the value (% power)

– a single press shows the

current setting. To alter either keep pressed or press again.
The value will then be stored without any further input.

Caution: Do not increase the power limit value to a value above the design level for the
oven or furnace model, or to a value above that correctly calculated for Silicon Carbide
elements. The heating elements could burn out, or other damage could be caused. Refer to
the main manual supplied with the Furnace or Oven.

3.5

Changing the Customer Calibration Type





Start at the Home display.
Press and hold the Page key for 1.5 seconds to enter the set up menu.
Press the Page key until

CL.St shows on the display.

Press the Up or Down arrow key to display the current calibration type -

(

FAct, C.CL1 or C.CL2)

Press the Up or Down arrow key to display the password screen.
Press the Up or Down arrow keys to enter the Calibration Password (Section 3.6).
Press the Page key to confirm the password.
Use the Up and Down arrow keys to select the calibration type.
The value will then be stored without any further input.

See Section 8 for more information on Customer Calibration.

3.6

Calibration Password


Once entered the calibration password remains active for 30 seconds after leaving the
setup menu to allow time to revisit if necessary.

The Calibration Password for this instrument is:

525

OPHi

Hold

CL.St

PASS

Hold