Customer Calibration

16

MC16-GB-C

– 1.04

8.2.1 Changing the Single Point Calibration Offset

OFSt






Starting from

CL.St after C.CL1 has been selected.

Press the Page key until

OFSt shows on the display.

Press the Up or Down arrow key to display the current calibration offset.
Press the Up or Down arrow key to change the offset value.
The value will then be stored without any further input.

Once the

C.CL1 has been selected as the customer calibration type, it is possible to start

from the Home display and go to

OFS.t directly and enter the password at this point when

the calibration adjustment is required again.

8.3

Dual Point Calibration -

C.CL2


Dual point calibration uses two offset values at two corresponding temperatures to
progressively change the calibration as the temperature increases or decreases. This is a
more accurate representation of how the temperature difference will occur.

Caution! - Do not make

CAL.L and CAL.H the same value as the controller will not work

correctly and could cause your furnace or oven to overheat.

8.4

Changing the Calibration, Low Temperature

CAL.L






Starting from

CL.St after C.CL2 has been selected.

Press the Page key until

CAL.L shows on the display.

Press the Up or Down arrow key to display the current calibration low temperature.
Press the Up or Down arrow key to change the temperature value.
The value will then be stored without any further input.

Once

C.CL2 has been selected as the customer calibration type, it is possible to start from

the Home display and go to

CAL.L directly (or any of the other settings in C.CL2) and enter

the password at this point when calibration adjustment is required again.

CL.St OFSt

Hold

If starting from Home Display

If password required

CL.St CAL.L

Hold

If starting from Home Display

If password required