Changing the calibration, high temperature, Cal.h – Carbolite 301 User Manual

Page 17

Advertising
background image

Customer Calibration

17

MC16-GB-C

– 1.04

8.5

Changing the Calibration, Low Temperature Offset -

OFS.L





OFS.L follows on from CAL.L when the Page key is pressed.
Press the Up or Down arrow key to display the current calibration low offset.
Press the Up or Down arrow key to alter the low offset value.
The value will then be stored without any further input.

8.6

Changing the Calibration, High Temperature

CAL.H






CAL.H follows on from OFS.L when the Page key is pressed.
Press the Up or Down arrow key to display the current calibration high temperature.
Press the Up or Down arrow key to change the temperature value.
The value will then be stored without any further input.

8.7

Changing the Calibration, High Temperature Offset -

OFS.H





OFS.H follows on from CAL.H when the Page key is pressed.
Press the Up or Down arrow key to display the current calibration high offset.
Press the Up or Down arrow key to alter the high offset value.
The value will then be stored without any further input.

OFS.L CAL.H

Hold

If starting from Home Display

If password required

CAL.L OFS.L

Hold

If starting from Home Display

If password required

CAL.H OFS.H

Hold

If starting from Home Display

If password required