High limit instrument setup parameters – Despatch LBB2-xxA/B-2 User Manual

Page 38

Advertising
background image

O

PERATION

LBB Series Oven Owner’s Manual

38

Version

18


Set High Limit instrument parameters using the Operating and Set-up modes (Section 5.4.2.1).
See section 5.4.4. to calibrate High Limit, only use SHif parameter. In most applications, it is not
necessary to alter the oven settings. The following instructions describe how to access, view and,
if desired, change the parameters.

If the

on the High Limit instrument is lit, press and hold

for four (4) seconds to enable

and . The High Limit instrument will automatically exit Setup mode if no keys are pressed

for about two (2) minutes.
1.

Press and hold

for four (4) seconds to enter Setup mode.

2.

Press

until desired parameter displays. See the Setup Parameter Table for more

information.
a.

The display will alternate between the parameter name and value.

b.

Use or to move to the desired setting.

c.

Press

to the value and advance to the next parameter.

3.

To leave setup parameters press

. The High Limit instrument automatically exits Setup

mode if no keys are pressed for about two (2) minutes.


5.4.7. High Limit Instrument Setup Parameters

Table 6 explains the High Limit instrument setup parameters.


When changing between centigrade and Fahrenheit, the setup
parameters
Filt, o1.Hy, HSP.L and HSP.H settings convert
automatically.


Table 6. High Limit instrument Setup Parameters.

Code

Name

Description

Settings

inPt

Input type

Select thermocouple type. LBB Series ovens use type
J thermocouples

J_tC

unit

Process unit.

Select between °C and °F for reading process
temperature

°

C (or

°

F)

rESo

Display
resolution.

Select location of decimal point on process-related
parameters

No.dP

SHif

PV shift value
(offset).

Moves the display temperature to the oven
temperature

0

Filt

PV filter.

If the process value is unstable to read, increasing this
value steadies the input signal.

0

out1

Output 1
function.

Function of the output. Must be set to Hi. for High Limit
control.

Hi.

o1.Hy

Output 1
hysteresis
value.

Amount of degrees the temperature must be below the
setpoint temperature before the High Limit can be reset 2.0