LBB Series Oven Owner’s Manual

M

AINTENANCE

Version

18

47

6.5.2. Replace High Limit Instrument

Danger!

Do not attempt any service on this oven before opening the
main power disconnect switch or removing all power from
the oven.

Tools needed:

One-quarter (1/4) inch socket set

Screwdriver

T20 Torx bit driver

1.

Disconnect power.

2.

Remove four screws from the face of the control panel (Figure 13).

3.

Slide the panel forward (Figure 14).

4.

Remove wires from the old High Limit instrument, carefully noting which wires connect to
which terminals (Figure 15).

5.

Disconnect High Limit instrument connector from circuit board (Figure 18).



6.

Disconnect the thermocouple wires (Figure 15).

7.

Press and hold down the tabs holding the High Limit instrument to the control panel and slide
it out (Figure 19).

High Limit
Connector
Block

Figure 18. High Limit Connector Block on Control Board.