Hi limit l91 hi-limit reset – Despatch LBB2-xxA/B-2 User Manual

Page 57

Advertising
background image

LBB Series Oven Owner’s Manual

T

ROUBLESHOOTING

Version

18

57


Operating Parameter Outline

Code

Name

Settings

CtL

Control Mode

PID

At

Auto-tuning

OFF

P *

Proportional Band

5 **

I

Integral Time

15

d

Derivative Time

OFF

Ct

Cycle Time

1

FL

Input Filter

2

bS

PV Bias (Offset)

0

LoC

Key Lock

0

* If P is not displayed the CtL must be first set to PID.
** For F change to value 13.

7.2. Hi Limit L91 Hi-Limit Reset

a.

If the DES2000 Output LED is ON and the oven is still not heating, is the OP1 LED
ON?. If it is, then the hi-limit is tripped and needs to be reset or the parameters need to
be verified.

b.

Press RESET key for several seconds until the LOCK LED goes out.

c.

If the OP1 LED is still on, press key for several seconds to enter the parameter mode.

d.

The INPT parameter should be displayed and alternating with J_T/C. If this parameter is
reading anything else other than J_T/C, press the down arrow until it does.

e.

Press RESET key for several seconds until the temperature (PV) is displayed.

f.

The PV reading is the actual oven temperature and HSP1 is the hi-limt alarm temperature
set point. This must be set 10 degrees greater than the DES2000 set point (SP) value.