Det-Tronics X5200 UVIR Flame Detector User Manual

Page 8

Advertising
background image

95-8546

6

13.1

ImPORTaNT SafETy NOTES

WarnIng

Do not open the detector assembly in a hazardous

area when power is applied. The detector contains

limited serviceable components and should never

be opened. Doing so could disturb critical optical

alignment and calibration parameters, possibly

causing serious damage.

CaUtIon

The wiring procedures in this manual are intended

to ensure proper functioning of the device under

normal conditions. However, because of the

many variations in wiring codes and regulations,

total compliance to these ordinances cannot be

guaranteed. Be certain that all wiring complies

with the NEC as well as all local ordinances. If

in doubt, consult the authority having jurisdiction

before wiring the system. Installation must be

done by a properly trained person.

CaUtIon

To prevent unwanted actuation or alarm,

extinguishing devices must be disabled prior to

performing detection system tests or maintenance.

CaUtIon

The UVIR flame detectors are to be installed in
places where the risk of mechanical damage is low.

attEntIon

Remove the protective cap from the front of the

detector before activating the system.

attEntIon

Observe precautions for handling electrostatic

sensitive devices.

attEntIon

The source tube is a flame-sealed gas tube

containing Neon, Hydrogen, and a trace amount
of Krypton 85 (Kr

85

), a radioactive material. The

total volume of gas within the tube is 0.6 ml per
tube, making the gas mixture inside the tube
nonflammable. If the gas envelope is broken, it
will not produce a flammable mixture, and the gas
immediately disperses into the air and is unlikely
to present any type of hazard. Krypton gas and its
radioactive isotope are inert and are not absorbed
by the body. No special handling measure or
personal protection equipment is needed for the
UVIR detectors.

The following is a partial list of compounds that exhibit
significant UV absorption characteristics . These are
also usually hazardous vapors . While generally of little
consequence in small amounts, these gases can restrict
UV detection if they are in the atmosphere in heavy
concentrations . It should also be determined whether or
not large amounts of these gases may be released as a
result of a fire-causing occurrence .

acetaldehyde

methyl methacrylate

acetone alpha-methylstyrene
acrylonitrile Naphthalene
Ethyl acrylate

Nitroethane

methyl acrylate

Nitrobenzene

Ethanol Nitromethane
ammonia 1-Nitropropane
aniline 2-Nitropropane
benzene 2-Pentanone
1,3 butadiene

Phenol

2—butanone Pyridine
butylamine

hydrogen Sulfide

Chlorobenzene Styrene
1-Chloro-1-Nitropropane Tetrachloroethylene
Chloroprene Toluene
Cumene Trichloroethylene
Cyclopentadiene

Vinyl Toluene

O-Dichlorobenzene Xylene
P-Dichlorobenzene

If UV-absorbing gases may be a factor in a given
application, precautionary measures should be taken .
Detectors can be placed closer to the potential hazard
area, and/or the sensitivity of the detection system can be
increased . Contact the factory for further details .

Substances such as methane, propane, butane, hexane,
camphor and octane are not UV absorbing .

absorption of infrared radiation in the range of 4 .2 to 4 .7
microns is not a significant problem with most organic
vapors, with the exception of those compounds that
have triple bonds such as acetylene, nitriles, silane, or
isocyanates . Carbon dioxide concentrations higher than
normally present in the atmosphere can also cause
substantial loss of fire detection sensitivity .

Table 3—UV and IR absorbing gases and Vapors

Advertising