Det-Tronics TXL Series Gas Transmitter User Manual

Page 4

Advertising
background image

CONNECTIONS—
Screw terminal.

CALIBRATION—
1-person, non-intrusive magnetic switch on cover.

ENCLOSURE—
NEMA 4, epoxy-coated aluminum enclosure.

SENSOR HOUSING—
Aluminum standard (stainless steel option available).

DIMENSIONS—
4.65" W x 4.5" H x 2.2" D.

SHIPPING WEIGHT—
1 lb (0.5 kg).

CERTIFICATION—
UL approved when properly installed for:

Class I, Groups A, B, C and D;
Class II, Groups E, F and G;
Class III.

CE and DEMKO approvals.

IS BARRIER—
Voc ≤ Vmax = 30 vdc,
Isc ≤ Imax = 125 ma,
Ca ≥ Ci + C cable,
La ≥ Li + L cable.

Suggested barrier models 9002/13-280/110/00 or
9002/13-252/121/04 manufactured by R. Stahl, Inc.

MODEL TXL-H2S

MEASUREMENT RANGE—
0-500 PPM standard (other ranges are field selectable).

ACCURACY—
±2 PPM or ±10% of applied gas concentration.

LINEARITY—
±5%.

REPEATABILITY—
±5%.

RESPONSE TIME—
T90 < 20 seconds.

SENSOR CROSS SENSITIVITY—
See Table 1.

TEMPERATURE RANGE—
–20°C to +50°C (–4°F to +122°F).

HUMIDITY RANGE—
15 to 90% (non-condensing).

WARRANTY—
2 years.

2

95-8541

1.1

Gas

Level

Reading

Carbon Monoxide

50 ppm

0

Hydrogen Sulfide

10 ppm

10

Sulfur Dioxide

2 ppm

0

Nitrogen Dioxide

3 ppm

0

Nitric Oxide

25 ppm

0

Chlorine

0.5 ppm

0

Hydrogen

100 ppm

0

Ethylene

100 ppm

0

Carbon Dioxide

5000 ppm

0

Ammonia

50 ppm

0

Table 1—Cross Sensitivity of H

2

S Sensor

Advertising