Det-Tronics TXL Series Gas Transmitter User Manual
Page 8
MODEL TXL-CL2
MEASUREMENT RANGE—
0-20 PPM standard (other ranges are field selectable).
ACCURACY—
±0.2 PPM or ±10% of applied gas concentration.
LINEARITY—
>10% full scale.
REPEATABILITY—
<±3%.
RESPONSE TIME—
T90 < 20 seconds.
SENSOR CROSS SENSITIVITY—
See Table 8.
TEMPERATURE RANGE—
–20°C to +40°C (–4°F to +104°F).
HUMIDITY RANGE—
15 to 90% (non-condensing).
WARRANTY—
18 months.
MODEL TXL-CLO2
MEASUREMENT RANGE—
0-9.9 PPM standard (other ranges are field selectable).
ACCURACY—
±0.2 PPM or ±10% of applied gas concentration.
LINEARITY—
>10% full scale.
REPEATABILITY—
<±3%.
RESPONSE TIME—
T90 < 20 seconds.
SENSOR CROSS SENSITIVITY—
See Table 9.
TEMPERATURE RANGE—
–15°C to +40°C (+5°F to +104°F).
HUMIDITY RANGE—
15 to 90% (non-condensing).
WARRANTY—
18 months.
6
95-8541
1.1
Gas
Level
Reading
Hydrogen Sulfide
25 ppm
–16 ppm
Nitrogen Dioxide
50 ppm
1.2 ppm
Hydrogen Chloride
9 ppm
1.2 ppm
Sulfur Dioxide
50 ppm
9 ppm
Carbon Dioxide
2000 ppm
0
Ammonia
50 ppm
–2 ppm
Table 8—Cross Sensitivity of Cl
2
Sensor
Gas
Level
Reading
Carbon Monoxide
1000 ppm
0
Carbon Dioxide
5000 ppm
0
Hydrochloric Acid
5 ppm
0
Chlorine
5 ppm
–1
Hydrogen
1%
0
Hydrogen Sulfide
10 ppm
0
Nitrogen
100%
0
Hydrogen Fluoride
3 ppm
0
Alcohols
1000 ppm
0
Table 9—Cross Sensitivity of ClO
2
Sensor