Environmental Express AutoBlock III User Manual

Page 15

Advertising
background image

START UP INSTRUCTIONS


Sample Injection Tubing / Pump Calibration

Follow the steps to Calibrate Sample Injection Tubing and Calibrate Pump under the Service Tab in this
manual.

The Sample Injection Tubing and Pump(s) has been tested and calibrated during manufacturing.
It is best to calibrate the system again to ensure accuracy.





Maintenance


Review and set up the Maintenance portion under the Service Tab in this manual.
The maintenance schedule has been set during manufacturing. There will be popup alerts on the touch
screen to remind the operator to inspect & maintain the AutoBlock.



Sensor Validation

Change of status may take a few moments. It is not an instant response.
If these sensors indicators do not change status please review the System Overview section of this
manual for setting them to a positive operating condition.

1. From the Manual Mode, turn the fan on by pressing the FAN button.
The Pressure Sensor indicator

P

on the touch screen, will turn green when operating correctly. The

front door must be closed.
2. Close the front door and the pressure sensor indicator

P

will turn red

3. Press the FAN button to turn the fan off. The pressure sensor indicator

P

will turn red.


The Leak Detector Sensor indicator

L

on the touch screen will be green when no leak is detected.







Manual Tab

You are now ready to successfully run digestion methods in your new AutoBlock.

Reference Program Tab for creating and editing methods.






ENVIRONMENTAL EXPRESS

For technical support call 1-800-745-8218 or send email to [email protected]

5528STARTUP rev 3

15

9/10/09