SERVICE TAB

The SERVICE TAB is set up for AutoBlock to be calibrated,

maintained, and custom levels set for optimal performance.







service tab screen

Setup Reagents

Select SETUP REAGENTS button from the SERVICE TAB

Reagents are set to correspond with reagent ports. AutoBlock reagent ports are located on the
left side of AutoBlock, labeled #’s 1-10. The reagents have been set up in the following manner:

Single System

Dual System**

*Port #1

H20(1)

*Port #6

H20(6)

Port #2

HNO3(2)

Port #7

KMnO4(7)

Port #3

HCL(3)

Port #8

PERSUL(8)

Port #4

H2O2(4)

Port #9

NH2OH(9)

Port #5

H2SO4(5)

Port #10

H20(10)


*Ports #1 & #6 must remain H

2

O

**In a single system ports #6-10 have been plugged.


If the reagent arrangement or if other reagents are used, you may change the list.
In the following steps. Listed below are common reagents with viscosity ratings.

Reagent

Viscosity

H2O(1)

(ports 1 & 6, cannot be changed)

Low

HNO3(2) Nitric

Acid

Medium

HCL(3) Hydrochloric

Acid

Low

H2O2(4)

Hydrogen Peroxide 30%

Medium

H2SO4(5) Sulfuric

Acid

High


H2O(6)

(ports 1 & 6, cannot be changed)

Low

KMnO4(7) Potassium

Permanganate

Low

PERSUL(8) per sulfates

Low

NH2OH(9) Hydroxylamine

Low

H2O(10) Water

Low


Viscosity ratings control the pump speed to maximize efficiency and provide volume delivery accuracy.
Low Viscosity, fast pump speed, (reagent near the viscosity of water)
High Viscosity, slow pump speed