HYDAC HLB 1300 User Manual

Page 24

Advertising
background image

User Manual HYDACLab

®

23.10.2008

9

8. Technical specifications

Input data

Relative humidity

0 .. 100 % saturation

Temperature

-25 ..

+100 °C

Dielectric constant (ε

R

)

1 ..10

Operating pressure

< 50 bar

Pressure resistance

< 600 bar

Fluid flow velocity

< 5 m/s

Output data - Humidity measurement

Output signal

4 .. 20 mA

(corresponds to 0 .. 100 %)

Calibration accuracy

≤ ± 2 % FS max.

Accuracy

≤ ± 3 % FS typ. *

Output data - Temperature measurement

Output signal

4 .. 20 mA

(corresponds to -25 .. +100 °C)

Accuracy

≤ ± 3 % FS max.

Output data - Relative change in dielectric constant (ε

R

)

Output signal

12 mA ± 8 mA

(± 30 % of IV)

Accuracy

see below **

Switching output
Signal 1 (N/C)

PNP transistor switching output 0.5 A max.
switching level

≥ U

B

– 4 V

Preset warning level SP1 relative humidity

≥ 85 %

Present warning level SP1 temperature

≥ 80 °C

Present warning level SP1 dielectric constant

± 15 %

(temperature compensated)

Ambient conditions

Nominal temperature range

+20 .. +80 °C

Storage temperature

-40 .. +90 °C

Fluid compatibility

Mineral oils HLP (HLP-D on request)
Esters: HEES, HETG
Seals: FPM

mark

EN 61000-6-1, EN 61000-6-2
EN 61000-6-3, EN 61000-6-4

Protection class to DIN 40050

IP 67

Other data

Supply voltage U

B

10 .. 36 V DC

Supply voltage residual ripple

≤ 5 %

Mechanical connection

G ¾ DIN 3852 E

Torque value

30 Nm

Electrical connection

M12x1, 5 pole

Reverse polarity, short circuit protection

Provided

Housing Stainless

steel

Weight approx.

205

g

FS (Full Scale); refers to full dynamic range IV (Initial Value)

* The best accuracy achievable when measuring relative humidity is heavily dependent on the type of fluid or fluid

additive(s) involved. More precise information on this is available on request.

** The measurement accuracy of relative change in dielectric constant

varies according to the application, the types of fluids involved and the sensor's own calibration.

More detailed information on this is available on request.