14. Important tips at a glance


• For hydraulic applications, we recommend that the sensor is fitted in a vertical position

with the electrical connector pointing downwards.

• The sensor once fitted must be completely immersed in the fluid and the fluid must be

able to circulate freely through the sensor.

• When installing without the mounting block ZBM 21 or ZBM 23, make particularly sure

that the flow of fluid is not impeded by the installation. There must be a minimum
distance of 10 mm between the front end of the sensor head and the opposite wall of the
pipe/housing.

• Ensure there is constant flow and prevent the formation of air bubbles. With standing oil,

variations in measurement may increase.

• In principle the HYDACLab

®

must only be supplied with voltage if full flow is guaranteed.

In other words, disconnect the HYDACLab

®

from the supply voltage as soon as the

system is switched off, stops operating or if the flow is interrupted for a significant period.

• Ensure that the operating conditions are typical during the optimization phase (learning

phase) i.e. the whole range of conditions relevant to operation are included.

• The start of the learning phase begins as soon as the HYDACLab

®

is connected to the

voltage supply and as soon as the temperature exceeds 25°C for the first time.

• During the optimization phase, avoid mixing fluids and ageing fluids.
• In the event of oil change or serious alteration in the ambient conditions, the unit should

be reset.