Table 1. system alerts, Instrument status, System alerts and alarms – INFICON 2020ppbPRO Portable Photoionization Detector User Manual

Page 52

Advertising
background image

46

ISP-074-576-P1D

46

Instrument Status

The instrument status is shown on the left of the first line of the status display. Each
status has a priority assigned to it. If more than one status is in effect, then the
status with the highest priority is displayed until the condition is corrected or until the
option is turned off. Table 1 is a list of the possible system alerts.

Table 1. System Alerts

2020ppbPRO
Display

Description

Zero Air Error

Zero gas too high

Span Gas Error

Span gas too low

Pump Error

Pump fault

UV Lamp Error

UV lamp fault

Blocked Filter

Blocked sample inlet filter

Low Battery Icon

Low battery

Data Log Full

Data memory full

The Unit is
Communicating with
the PC

Instrument communicating with PC

Instrument Over
Range

Sample concentration over instrument
operating range

Alarm + R Icon

Real time concentration alarm

Alarm + T Icon

TWA concentration alarm

Alarm + S Icon

STEL concentration alarm

System Alerts and Alarms

While operating the instrument, system alerts can occur. To accurately identify the
source of the alarm, each type of alarm has been given a unique status.

In addition to the status, the 2020ppbPRO also has an audio alarm and an alarm
LED. To conserve power, the 2020ppbPRO alternates between the LED and audio.
Different alarms are identified by the frequency at which the 2020ppbPRO alternates
between the audio and LED; Peak alarm is 5 times per second, STEL alarm is 2.5
times per second, and TWA alarm is 1.25 times per second.