90

ISP-074-576-P1D

90

Inlet filter:

Replaceable Teflon/Polypropylene

Inlet flow rate:

Greater than 300 mL/min

Operating
temperature
range:

0 to 50°C (32 to 122°F)


I/S Certified
Temperature
Range:

0 to 40

0

C (32 to 105

0

F)

Operating

0 to 95% relative humidity

humidity range:

(non-condensing)


Operating

0 to 100% relative humidity

humidity range

(non-condensing)

with optional
humidity pre-
filter tube

Operating
altitude
range:

Up to 10,000 ft (3000 m)


Storage
temperature
range:

-10 to 65°C (14 to 149°F)


Storage

humidity range:

0 to 100% relative humidity (non-condensing)

Storage
altitude
range:

Up to 15,000 ft (4570 m)

Operating

concentration
range:

10 ppb to 40 ppm, isobutylene

Accuracy:

+/-10% or +/-2 ppm, whichever is greater

Precision:

1% of calibration (calibrated with 10 ppm
isobutylene)

Response time:

Less than 3 seconds to 90%

LDL:

10 ppb isobutylene