To empty the data logger in the 2020ppbPRO:

1. Press the MENU key.

2.

Use the DOWN ARROW key to choose “Data Logger Options”, and then
press the SELECT key.

3.

Choose “Clear Data Log”, and then press the SELECT key.

4.

Press either the YES or NO key in response to “Are you sure you want to
clear all data?”.

“Download to PC” begins download of stored data from the 2020ppbPRO to the PC.
The 2020ppbPRO computer cable, MX750120, must be connected between the
2020ppbPRO and the PC before beginning the data download. The 2020ppbPRO
software package, the 2020ppbPRO Comm, must be installed and running on the
PC prior to download.

To download data:

1. Press the MENU key.

2.

Use the DOWN ARROW key to choose “Data Log Options” the press the
SELECT key.

3.

Use the DOWN ARROW key to choose “Download to PC”, then press the
SELECT key.

4.

Press the NEXT key at the “Connect Instrument to PC” prompt.

5. The display will show

“Downloading Data” and data will now download to

the PC. Press the DONE key to stop the download at any time.

Unit Setup Functions

Unit setup functions are used to select the 2020ppbPRO features. There are nine
functions which can be set on the 2020ppbPRO; Pump, Backlight, User Mode,
Clock, Date Format, Language, Units of Measure, and Sample Mode.
Figure
13
shows a menu detailing the User Setup functions. Press the MENU key in any
operating mode to access “Unit Setup”.