5. Preparations

5.1. Installation

Caution:

Securely fix the immersion circulator. The heater may not be in contact with the
wall of the bath tank. Keep a distance of at least 15 mm.
Units not adequately fixed may drop into the bath tank.
D a n g e r o f e l e c t r i c s h o c k !
First pull out the power plug to disconnect the unit from the power supply net.
Then take the immersion circulator out of the bath tank.
Make a service technician check the instrument before it is used again.

 Place the unit on an even surface on a pad made of non-

flammable material.

 The heating immersion circulator is mounted using a bath

attachment clamp (21) designed for bath wall thicknesses
up to 26 mm.

21

22

 Use the two sleeves (22) supplied with the unit to reduce

the immersion depth from 165 mm to 145 mm (see
drawing).

 For use with glass vessels an upright stand rod,

available as optional accessory (order no. 8 970 022)
may be attached.