JULABO MA Heating Immersion Circulator User Manual

Page 24

Advertising
background image

Setting of temperatures

24


Refer to page 32 for switch-over to 3-temperature-mode

CFG = CONFIGURATION

3SP = 3 SETPOINT

Setting of working temperature in the

menu

1. Press the key

. The value >tx<

flashes

2. Select SETPOINT >t 1< or >t 2< or >t 3< using the key

or

.

3. Confirm by pressing the

key.

The circulator uses the new working temperature value for
temperature control.

Example: setting / adjustment of of pre-settings of "t 3"


1. Press

the

key. The parameter >tx< flashes.

2. Select the setpoint >t3< by pressing

or

.

3. Keep the key

pressed until the integer digits flash

.

(example: <70>)

4. Change the value by pressing

and

to 85.00 °C and

confirm by pressing the

key.

The decimal digits flash

and can be adjusted if desired.

Confirm once more by pressing the

key.

Example on the left: SETPOINT >t3< / 85.00.
The end of the adjustment is signaled by the flashing message
>t3<

If the active setpoint (SETPNT) is changed, the new value is
immediately used for the control of the working temperature.
The heater control indicator flashes.

If the other two setpoints (not activated for control) are changed
the

MENU has to be left by pressing the

key after

the decimal digits have been confirmed

Notice: Refer to chapter
9.6. MENU LIMITS

Advertising