JULABO MA Heating Immersion Circulator User Manual

Page 47

Advertising
background image

MA

47

Command

Parameter

Response of circulator

in_sp_03 none High

temperature warning limit

„t High“

in_sp_04 none Low

temperature warning limit

„t Low“

in_sp_07

none

Pump pressure stage

in_par_01

none

Te - Time constant of the external bath.

in_par_02

none

Si - Internal slope

in_par_03

none

Ti - Time constant of the internal bath.

in_par_06

none

Xp control parameter of the internal controller.

in_par_07

none

Tn control parameter of the internal controller.

in_par_08

none

Tv control parameter of the internal controller.

in_mode_01 none

Selected

setpoint:

0 = Setpoint „t 1“
1 = Setpoint „t 2“
2 = Setpoint „t 3“

in_mode_05 none

Circulator

in Stop/Start condition:

0 = Stop
1 = Start


12.4. Status

messages

Status messages

Description

00 MANUAL STOP

Circulator in „OFF“ state.

01 MANUAL START

Circulator in keypad control mode.

02 REMOTE STOP

Circulator in „r OFF“ state.

03 REMOTE START

Circulator in remote control mode.


12.5. Error

messages

ERROR MESSAGES

Description

-01 LOW LEVEL ALARM

Low liquid level alarm.

-02 REFRIGERATOR ALARM

Control cable of the refrigerated circulator or
MVS solenoid valve controller short-circuited
or interrupted.

-03 EXCESS TEMPERATURE WARNING

High temperature warning.

-04 LOW TEMPERATURE WARNING

Low temperature warning.

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or
interrupted.