Remote control

48

ERROR MESSAGES

Description

-06 SENSOR DIFFERENCE ALARM

Sensor difference alarm.
Working temperature and safety sensors
report a temperature difference of more than
35 K.

-07 I2C-BUS ERROR

Internal error when reading or writing the I2C
bus.

-08 INVALID COMMAND

Invalid command.

-09 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE

Invalid command in current operating mode.

-10 VALUE TOO SMALL

Entered value too small.

-11 VALUE TOO LARGE

Entered value too large.

-12 TEMPERATURE MEASUREMENT
ALARM

Error in A/D converter.

-13 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS

Value lies outside the adjusted range for the
high and low temperature warning limits. But
value is stored.

-14 EXCESS TEMPERATURE
PROTECTOR ALARM

Excess temperature protection alarm

-20 WARNING: CLEAN CONDENSOR OR
CHECK COOLING WATER CIRCUIT OF
REFRIGERATOR

Cooling of the condenser is affected.
Clean air-cooled condenser.
Check the flow rate and cooling water
temperature on water-cooled condenser.

-21 WARNING: COMPRESSOR STAGE 1
DOES NOT WORK

Compressor stage 1 does not work.

-22 WARNING: COMPRESSOR STAGE 2
DOES NOT WORK

Compressor stage 2 does not work.

-23 WARNING: HIGH TEMPERATURE
ON COMPRESSOR STAGE 1

Excess temperature on compressor stage 1.

-24 WARNING: HIGH TEMPERATURE
ON COMPRESSOR STAGE 2

Excess temperature on compressor stage 2.

-25 REFRIGERATOR WARNING

Error in the refrigerating machine.

-30 CONFIGURATION ERROR:
CONFIRM BY PRESSING <ENTER> ON
CIRCULATOR

The configuration of the circulator does not
conform to its present use.

Press

to automatically perform a single

modification of the configuration.

-33 SAFETY SENSOR ALARM

Excess temperature sensor short-circuited or
interrupted.

-40 NIVEAU LEVEL WARNUNG

Low liquid level warning in the internal
reservoir.