9.5.3. Example:

3-point calibration for internal control

In the temperature range from 80 °C to 160 °C the calibration curve of
the temperature sensor (TT) is to be adjusted to the actual temperatures
at measuring point (CT).

80.00 °C

120.00 °C

160.00 °C

1. Set working temperature setpoint :

Refer to „Direct temperature setting “ page 23

1. By pressing the key

or

the circulator switches to the

active >SETPOINT< see example on the left: >t1 25.00°C<.

The integer digits flash

(Example: <25>).

2. Change the value to 80.00 °C by pressing the keys

and

and

confirm by pressing the key

.

The decimal digits flash.

Confirm once more by pressing the key

.

3. The bath is heated up.

Wait for approx. 5 minutes until the temperature is constant.

2. Reading of temperature measuring device

Read the value of measuring point CT on the device and enter under
menu item >Ct X< by using the keypad.

>Ct 1< (79.70 °C)
>Ct 2< (119.5 °C)
>Ct 3< (159.3 °C)

3. Calibration

Menu level 1

Press the

key if parameter is to be retained. Correction

function for parameters or values (prior to OK).

Setting is required only for the first calibration point.

Level 2

Parameter level

An ATC calibration is to be carried out.
Set to >no<

The parameter flashes, switch by pressing

and

.

 The parameter flashes, switch by pressing

and

.

A >3-point< calibration is carried out.