Temperature measurement range:
Forehead mode: 34~42.2°C (93.2~108°F)
Surface mode: -22~80°C (-7.6~176°F)
Operating temperature range: 10~40°C (50~104°F)
Storage temperature range: It should be stored at room temperature between -20~+50°C , RH≦85%

Transportation temperature shall be less than 70°C, RH≦95%

Comply with ASTM E1965-98, IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safety) standards.
Accuracy:
Forehead mode: ±0.2°C for 36.0~39.0°C, others ±0.3°C (0.5°F)
Surface mode: ±4% of reading or ±2°C (4°F) whichever is greater
Fever alarm, memory and °C/°F switch function
Battery: one lithium cell battery (CR2032 *1pcs). Battery life is estimated at 3,000 readings.
Enable normal use: 1 year
This thermometer converts the forehead temperature to display its `ORAL VALUE EQUIVALENT'.

There is no gender and age limitation for using the infrared thermometer.
This is not an AP or APG product.