Guidance and manufacturer’s declaration – electromagnetic immunity

The NCT1 is intended for use in the electromagnetic environment specified below. The customer or the user of the NCT1

should assure that it is used in such an environment.

Immunity test

Compliance

level

Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

Not

applicable

3 V/m

IEC 60601

test level

3 Vrms

150 kHz

to 80 MHz

3 V/m

80 MHz

to 2,5 GHz

Portable and mobile RF communications equipment should be used no closer

to any part of the NCT1, including cables, than the recommended separation

distance calculated from the equation applicable to the frequency of the

transmitter.
Recommended separation distance

where P is the maximum output power rating of the transmitter in watts (W)

according to the transmitter manufacturer and d is the recommended separation

distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic

site survey, a should be less than the compliance level in each frequency range b.
Interference may occur in the vicinity of equipment marked with the following

symbol:

80 MHz to 800 MHz

800 MHz to 2,5 GHz

d = 1,2

d = 1,2

d = 2,3