EMC Table

Guidance and manufacturer’s declaration – electromagnetic immunity

The NCT1 is intended for use in the electromagnetic environment specified below. The customer or the user of the

NCT1 should assure that it is used in such an environment.

Immunity test

Compliance level Electromagnetic environment – guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

Electrical fast

transient/burst

IEC 61000-4-4

6 kV contact

8 kV air

6 kV contact

8 kV air

Not applicable

Floors should be wood, concrete or ceramic tile. If floors are

covered with synthetic material, the relative humidity should

be at least 30 %.

Mains power quality should be that of a typical commercial

or hospital environment.

Mains power quality should be that of a typical commercial

or hospital environment.

Surge

IEC 61000-4-5

Not applicable

IEC 60601

test level

2 kV for power

supply lines

1 kV for input/

output lines

1 kV line(s) to

line(s)

2 kV line(s) to

earth