EMC Table

Guidance and manufacturer’s declaration – electromagnetic immunity

Immunity test

Compliance level Electromagnetic environment – guidance

interruptions

and voltage

variations on

power supply

input lines

IEC 61000-4-11

<5 % UT

(>95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles

<5 % UT

(>95 % dip in UT)

for 5 sec

Mains power quality should be that of a typical commercial

or hospital environment. If the user of the NCT1 requires

continued operation during power mains interruptions, it is

recommended that the NCT1 be powered from an

uninterruptible power supply or a battery.

Power frequency magnetic fields should be at levels

characteristic of a typical location in a typical commercial

or hospital environment.

Power frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.

Not applicable

IEC 60601

test level

3 A/m

3 A/m