2 operation – Labnet Z206A High Capacity, Compact Research Centrifuge User Manual

Page 16

Advertising
background image

Page 16 of 24

INSTRUCTION MANUAL Z 206 A

© HERMLE Labortechnik GmbH

2 OPERATION

2.2.4 Pre-selection of speed / RCF-value

By pressing the key “rpm/rcf“ (6) the pre-selection of the speed is activated. By pressing the key once

the word “rpm“ (12) flashes.
By pressing the key once again the pre-selection of the centrifugal forces may be chosen. Flashing the

word “rcf“ (13).
You can set the desired values with the adjusting knob (7). In the display (14) the regulated value is
shown permanently, before, during and after the run.

The speed is adjustable between 200 rpm and maximum revolution of the centrifuge resp. the maximum
permissible revolution of the pre-selected rotor.
It is the same with the pre-selection of the RCF-value. The setting range is between 20 xg and the
maximum permissible centrifugal force of the rotor.
The maximum speed of the Z 206 A is 6000 rpm resp. 4180 xg.



Photo 13


Maximum speed of the rotors available for the Z 206 A

Rotor-Number Max.

Revolution

RCF

Value

220.96 V01

6000 rpm

4180 xg

220.97 V01

6000 rpm

3820 xg

220.95 V06

6000 rpm

2930 xg

220.68 V04

3500 rpm

1450 xg

Attention:
Please also check the maximum permissible speed of your test tubes.





14