INSTRUCTION MANUAL Z 206 A

© HERMLE Labortechnik GmbH

2 OPERATION

2.2.6 Pre-selection of brake intensity and acceleration


By pressing the key “accel/decel“ (5) this function is activated.

By pressing the key once the word “accel“ (16) flashes in the display “acc/dec“. The desired
acceleration can be pre-selected by the adjusting knob (7). The value 0 is equivalent to the lowest and

the value 9 to the highest acceleration.
By pressing the key “accel/decel“ (5) twice, in the display “acc/dec“ indicates the word “decel“ (17). Now
the desired brake intensity can be pre-selected by the adjusting knob (7). The value 9 is equivalent to

the shortest and the value 0 to longest possible brake time.



Photo 15


Acceleration and deceleration times Z 206 A(120 V / 230 V) in seconds

Acceleration

values

Deceleration

values

Rotor-Number

Level

0 Level

9 Level

0 Level

9

220.96

V01

70 40 150 35

220.97

V01

80 35 150 35

220.95

V06

60 11 30 10

220.68

V04

35 8 25 7