1 PRODUCT DESCRIPTION

1.1 Usage in accordance with safety standards

1.1.1 General

information


This unit is a medicine product according to the IVD guide line 98/79/EG.


1.1.1.1

Hazards

and

precautions


Before setting the centrifuge into operation, please read this instruction manual carefully!


For your personal safety, please review the following precautions:

• The HERMLE Z 206 A is not explosion-proof and must therefore not be operated in explosion-
endangered areas or locations. During centrifugation, it is prohibited to stay within the safety zone of

30 cm around the centrifuge or deposit hazardous substances within this area.

• Centrifugation of flammable, explosive and radioactive substances, which chemically
react with high energy, is strictly prohibited!

• Never spin toxic or pathogenic material without adequate safety precautions, i.e. centrifugation of
buckets / tubes without or with defective hermetic sealings is strictly prohibited.
The user is obligated to perform appropriate disinfection procedures in the case that dangerous
substances have contaminated the centrifuge and / or the accessories. When centrifuging infectious
substances, always pay attention to the General Laboratory Precautions. If necessary, contact your
safety officer.


• It is prohibited to run the centrifuge with rotors that are not listed for this unit.


• Under no circumstances open the lid of the centrifuge while the rotor is still running or rotating with a
speed of > 2 m/s.


Following rules must strictly be adhered to:

• Do not operate the centrifuge in the case that it is not installed correctly.

• Do not operate the centrifuge when dismounted (e.g. without housing).

• Do not run the centrifuge when mechanical or electrical components have been tampered with
by unauthorized personnel.

• Do not use accessories such as rotors and buckets, which are not exclusively approved by
HERMLE Labortechnik GmbH, except commercially available centrifuge tubes made of glass

or plastic.

• Do not spin extremely corrosive substances, as they may cause material damages and impair
mechanical resistance.


• Do not operate the centrifuge with rotors or buckets, which show any signs of corrosion or mechanical

damage.