5 installation site requirements, Installation site requirements, Instrument setup – Leica Biosystems LPC Operating Manual User Manual

Page 20

Advertising
background image

20

Operating Manual V 1.0 – 06/2010

4.5

Installation site requirements

The installation site must be well-ventilated.
Never operate the instrument in rooms with explosion hazard.

4.

Instrument Setup

The printer needs an installation area of approx. 260 x 460 mm and the fil-
ter unit needs an additional installation area of approx. 470 x 490 mm.
Additional space is required for the PC containing the control software.

Stable laboratory bench and practically vibration-free floor.

Room temperature consistently between +5 °C and +40 °C.

Relative air humidity maximum 80 %, non-condensing.

Avoid vibrations, bright direct light and strong temperature fluctuations.

The AC socket used for the power supply must be close to the laser
printer and easily accessible.

The laser printer is intended for use in enclosed rooms only and may be
used at elevations up to 2000 m only.