The prefilter is replaced by opening the prefilter compartment of the filter
unit and replacing the prefilter bag (17).

Open the prefilter compartment by pulling the tube end (18) of the plastic
cover (19).

1. After installing the new filter bag, first fasten the lower part of the cover

to the metal plate.

The lower side of the cover has 2 metal tabs that must be inserted into
the slots on the bottom of the plastic cover.

If the metal tabs are not inserted into the slots in the plastic
cover, the cover will break.

5.

Cleaning and Maintenance

17

18

19

Fig. 29

To replace the filter, turn it counterclockwise as
shown above and pull it off. Remove the filter
bag carefully and dispose of it according to the
regulations applicable in the country of use.
Fit the new filter bag over the Allen screws and
fasten it in place by clamping/turning clockwise.