5 replacing fuses, Replacing fuses, Cleaning and maintenance – Leica Biosystems LPC Operating Manual User Manual

Page 35

Advertising
background image

35

Leica LP C

5.5

Replacing fuses

Before replacing the secondary fuses, always switch the in-
strument off first and unplug it from the power supply.

Fig. 34

Fig. 35

The fuses are located on the rear of the instrument.

Using a screwdriver (40), pry out the fuse in-
sert (17) and remove it.

40

17

Remove the defective fuse from the fuse hold-
er (17) and replace it with a replacement fuse
(17a).

Fuse: F 5AL 250 V

Replace the fuse insert with the replacement
fuse until it clicks into place audibly.

17

17a

5.

Cleaning and Maintenance