If the Safety Interlock operates, the
user must:

Immediately investigate the reason for the in-
terlock.

Check the housing cover for damage before
starting up the Cassette Printer.

If damage is found or the instrument does not
work, stop using the Cassette Printer and
contact the Leica service technician or Sales
Consultant to have the problem remedied.

2. Mechanical design

The beam path from laser tube to final pass
through lens is entirely enclosed in a metal
shield, with suitable material that will absorb
the heat produced by the beam emitting from
the laser.

3. Password security access

The software requires the user to enter an
access password to change the layout of the
cassettes of the operating parameters of the
Cassette Printer itself, including the laser cy-
cle. The password-protected areas of the Cas-
sette Printer software may be accessed only
after consultation with the Leica service tech-
nician.

The features and precautions described above
are designed with the safety of the user in mind.
Should you have any questions or suggestions
please contact Leica directly.

24

Fig. 16

11 - Safety interlock pin

12 - Intake slot for safety

interlock pin

25

4.

Instrument Setup

Safety Interlock (24/25) in the main cover.