Safety Instructions

13

Safety Instructions

General Recommendations

Read the safety and operating instructions before operating the
display.

Retain safety and operating instructions for future reference.

Adhere to all warnings on the display and in the operating
instructions manual.

Follow all instructions for operation and use.

Electrical shock

Type of protection (electrical):

Display with external power supply: Class I equipment

Degree of safety (flammable anesthetic mixture):

Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.

Non-patient care equipment

Equipment primarily for use in a health care facility that is intended
for use where contact with a patient is unlikely.

Power connection - display with external power supply

• Power requirements: The display must be powered using the

delivered medical approved 24 VDC SELV power supply.