Choose 4-20 mA or 10-50 mA operation.

Works with SOURCE mA, Read mA and Simulate 2-Wire
Transmitters.

Move the power switch

w

to SOURCE or READ and Double

click the

e

DIAL KNOB and the MAIN menu for the function in

use will appear for 15 seconds:

Turn the

e

DIAL KNOB to move the indicator to

FUNCTION and

press the DIAL KNOB until

mA appears. Turn the

e

DIAL KNOB

to move to the second menu page so the word

mA RANGE

appears at the top of the menu.

Turn the

e

DIAL KNOB to move through the menu. Press the

e

DIAL KNOB to toggle between

4-20mA and 10-50mA or to turn

on

LOOPSCOPE (2-Wire Transmitter only) or LEAK DETECT

(POWER MEASURE only.

mA RANGE

> EXIT

(2/3)

mA RANGE

4-20mA

10-50mA

LOOPSCOPE OFF

ON

Operating Instructions

MAIN

> EXIT

(1/3)

FUNCTION

mA

MODE

SOURCE

UNITS

mA

HART 250

Ω* ON

mA RANGE

> EXIT

(2/3)

mA RANGE

4-20mA

10-50mA

LEAK DETECT OFF

ON

POWER
MEASURE
MENU

2-WIRE
SIMULATE
MENU

* Hart 250Ω only appears

in 4-20 mA operation