Pr eliminar y – Practical Instrument Electronics 850 User Manual

Page 16

Advertising
background image

Page 14

Pr

eliminar

y

2 Wire SIM mA, 2 Wire SIM % (Percent of 4-20 or 10-50 mA)

Choose this function to simulate a 2 Wire Transmitter output from
0.000 to 24.000/0.000 to 52.000 milliamps. Operates in loops
with power supply voltages from 2 to 60 VDC.
Move the power switch

w

to SOURCE then Double Click the

EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select mA for the FUNCTION and 2W SIM for the MODE.
Choose either mA or % and whether you need the 250Ω HART

resistor active in the loop (250Ω

resistor is only available in

4-20mA operation).

If you would also like to see the condition of the loop turn the

e

DIAL KNOB to move to the second (or third) menu page so the
word

mA RANGE appears at the top of the menu.

LOOPSCOPE turns on the Loop Diagnostic display which
simultaneously indicates the current, voltage and resistance of
the loop. When enabled the

LOOPSCOPE also indicates errors

by flashing the LED and writing diagnostic messages on the
display. Messages include

LOW LOOP VOLT, HIGH LOOP

VOLT and HIGH LOOP IMPED. With LOOPSCOPE on the 850
occasionally performs a test on the loop. During this short test
the loop mA signal will be affected.

2-WIRE

SUPPLY

25.4 V

LOAD

485 Ω

12.000

mA

2-WIRE

12.000

mA

850 with LOOPSCOPE

850 Simulating a

Transmitter

mA RANGE

> EXIT

mA RANGE

4-20mA

10-50mA

LOOPSCOPE OFF

ON

Advertising