Pr eliminar y – Practical Instrument Electronics 850 User Manual

Page 27

Advertising
background image

Page 25

Pr

eliminar

y

Thermocouple Source

Choose this function to provide a simulated thermocouple signal
into controllers, temperature transmitters, indicators or any input
devices that measure thermocouple sensors.

Move the power switch

w

to SOURCE then Double Click the

EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select T/C for the FUNCTION, °F or °C for the UNITS, T/C Type
(J, K, E, T, R, S, B, N, L (J-DIN), U (T-DIN), G, C, D or P (Platinel
II)) and internal COLD JUNC ON or OFF (ON is the default).

Connect the PIE 850 to the inputs of the device being calibrated
using the proper type of thermocouple wire via the miniature
thermocouple socket.

Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in 0.1°
increments by turning the knob

e

. Press and turn the knob for

faster dialing with 10.0° increments.

T/C

All mA∙ Read V

-

2

3

RTD

RTD

6

5

4

+

Source & Read Hz∙ mV

Source V∙ pH

-

+

1

-

All thermocouple connections

Instrument with T/C Input

Controller

Temperature Transmitter

Temperature Indicator

Temperature Trip or Alarm

Thermocouple Wire

HI

1083.0

°C

T/C

+

-

TYPE K

Advertising