Pr eliminar y – Practical Instrument Electronics 850 User Manual

Page 7

Advertising
background image

Page 5

Pr

eliminar

y

Operating Instructions

STORING HI and LO EZ-CHECK Source Outputs

Speed up your calibration by storing Span & Zero output setting
for instant recall with the EZ-CHECK switch.

1) Store your high (SPAN) output temperature by moving the

EZ-CHECK switch to the

HI position and turning the EZ-Dial

knob until the desired output value is on the display. Press
and hold the EZ-Dial knob until

STORED appears to store

the value. Release the EZ-Dial knob.

2) Store your low (ZERO) output value by moving the EZ-CHECK

switch to the

LO position and turning the EZ-Dial knob until

the desired output value is on the display. Press and hold the
EZ-Dial knob until

STORED appears to store the value.

Release the EZ-Dial knob.

3) Instantly output your SPAN and ZERO output settings by

moving the EZ-CHECK switch between HI and LO. You may
also select any third output setting (such as mid-range) using
the SET position on the EZ-CHECK switch.

Connections
PIE 850 has banana jacks compatible with unshielded or
retractable banana plugs. Included with your calibrator are a pair
of test leads with alligator clips for mA, V, pH & Hz connections.
Four test leads with spade lugs are also included for 2, 3 and 4
Wire RTD connections. Thermocouple connections are made
through a miniature thermocouple socket.

T/C

All mA∙ Read V

-

2

3

RTD

RTD

6

5

4

+

Source & Read Hz∙ mV

Source V∙ pH

-

+

1

-