Read mA
Ranges and Resolution 0.000 to 24.000 mA or -25.00 to 125.00%

of 4-20 mA, 0.000 to 52.000 mA or -25.00

to 105.00% of 10-50 mA

Accuracy

≤ ± (0.02 % of Reading + 0.003 mA)

Voltage burden

≤ 2V at 24 mA or 52 mA.

Overload/Current

limit protection

25 mA nominal

Source mA / Power & Measure Two Wire Transmitters
Ranges and Resolution

Same as Read mA

Accuracy

≤ ± (0.02 % of Reading + 0.003 mA)

Loop compliance voltage ≥ 24 DCV @ 20 mA, ≥ 40 DCV @ 52 mA
Loop drive capability

1200 Ω at 20 mA for 15 hours nominal;

800 Ω at 50 mA for ## hours nominal

Subtract 250 Ω with Hart Resistor

enabled (4-20 only)

mA 2-Wire Transmitter Simulation
Accuracy

Same as Source/Power & Measure

Voltage burden

≤ 2V at 20 mA or 52 mA

Overload/Current limit 24 mA nominal in 20 mA range, 53 mA

nominal in 50 mA range

Loop voltage limits

2 to 60 VDC (fuse-less protected)

Specifications

DC Voltage Read
Range and Resolution

±99.999 mV, ±999.99mV, 0 to 10.250 V, 0.00

to 60.00 V DC

Accuracy

≤ ± (0.02 % of Reading + 0.01% Full Scale)

Input resistance

≥ 1 MΩ