Pr eliminar y – Practical Instrument Electronics 850 User Manual

Page 19

Advertising
background image

Page 17

Pr

eliminar

y

Milliamp Output Signal

Controller

Transmitter

P/I

DCS

MAX

19.999

mA

READ

5 6

+ -

READ mA, READ % (Percent of 4-20 or 10-50 mA)

Choose this function to measure from 0.000 to 24.000/0.000 to
52.000 milliamps or -25.00 to 125.00%.

Move the power switch

w

to READ then Double Click the

EZ-DIAL knob to get into the Menu. Turn the knob

e

to scroll

through the settings and press the knob to make your selection.
Select mA for the FUNCTION and READ for the MODE. Choose
either mA or % and whether you need the 250

HART resistor

active in the loop (250Ω

resistor is only available in 4-20mA

operation).

Connect the red input lead (+) of the PIE 850 to the more
positive point of the break and the black input to the more
negative point.

Signals below 0 mA or open circuits are indicated by 0.000 mA
(-25.00%) on the display. Signals above 24 or 52 mA are current
limited by protection circuitry with “OVERRANGE” flashed on the
display and the red OVERLOAD LED lit.

The PIE 850 measures the input signal and constantly updates
the display with the current reading. Move the EZ-CHECK switch

q

to MAX to see the highest reading and to MIN to see the

lowest reading. Press and hold the knob

e

to clear the MAX and

MIN readings.

T/C

All mA∙ Read V

-

2

3

RTD

RTD

6

5

4

+

Source & Read Hz∙ mV

Source V∙ pH

-

+

1

-

All milliamp connections